Instrument - Scanning Mobility Particle Sizer
Short name:
SMPS
Full name:
Scanning Mobility Particle Sizer
What is being measured:
Sampling Protocol:
Online
Manufacturer:
Home made
Model:
TSI3081, TSI3771
Instrument year
:
2005
Data recording software:
Labview
Data analysis software:
Custom Excel Program
Raw data time resolution:
83s
Analysis data averaging:
None
Detection limit:
#: 0.108 cm-3; V: 0.001 um3/cm3
Sensitivity to temperature (and correction method, if applicable):
:
Up to 0.5% in # or 9.5% in Volume between 5C and 35C
Correction Method: Calculate different inversion matrices for different temperatures and apply to the raw data.
Sensitivity to relative humidity (and correction method, if applicable):
:
N/A
Sampling method:
Direct gas sampling
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
12
Length of tubing (cm):
0.03
Instrument flow rate:
2.58
Tubing inner diameter:
Tubing material:
Copper
Chemical identification method:
None
Data analysis method:
The SMPS data that were recorded in scanning mode are analyzed using a custom Excel script. Particle size distribution, number/volume concentration as a function of time are displayed. Custom wall loss fitting time is available to do wall loss correction as needed.
Quantification method:
Signal strength is converted to partial number/size distribution by designed transfer functions.
Calibration method:
Standard particles(created by standard PSL solution) with a certain size range are sampled and detected. The calibration factor will be calculated based on the difference.
Calibration drift estimate:
Low (on the order of weeks)
Calibration schedule:
Annually
Uncertainty estimation method:
Duplication of the same experimental conditions.
Known interferences:
N/A
Link to supplemental information:
Additional notes:
Measurement uncertainty:
Measurement units: