Instrument - Aerosol Mass Spetrometer
Short name:
AMS
Full name:
Aerosol Mass Spetrometer
What is being measured:
Sampling Protocol:
Online
Manufacturer:
Aerodyne
Model:
HR-ToF-AMS
Instrument year
:
None specified
Data recording software:
ToF-DAQ
Data analysis software:
Igor
Raw data time resolution:
150s
Analysis data averaging:
None
Detection limit:
3 ng/m3 for V mode; 32 ng/m3 for W mode
Sensitivity to temperature (and correction method, if applicable):
:
N/A
Sensitivity to relative humidity (and correction method, if applicable):
:
N/A
Sampling method:
Direct gas sampling
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
0.01
Instrument flow rate:
0.08
Tubing inner diameter:
Tubing material:
Copper
Chemical identification method:
EI fragmentation
Data analysis method:
The SMPS data that were recorded in scanning mode are analyzed using a custom Excel script. Particle size distribution, number/volume concentration as a function of time are displayed. Custom wall loss fitting time is available to do wall loss correction as needed.
Quantification method:
For V mode, ToF calculation;
For W mode, the software dissects the measured peaks to several peaks of known compounds around there (precision to 0.0001 m/z) and does the statistics with time or run stamps.
Calibration method:
Size-selected NH4NO3 particles are measured using two modes. ToF Ion Efficiency factors will be calculated based on the comparison between the data from SMPS and AMS.
Calibration drift estimate:
Low (on the order of weeks)
Calibration schedule:
Monthly
Uncertainty estimation method:
Duplication of the same experimental conditions.
Known interferences:
CO2
Link to supplemental information:
Additional notes:
Measurement uncertainty:
Measurement units: