Instrument - Chemiluminescence NO instrument (Weinheimer/Montzka/Tyndall)
Short name:
NO-CL
Full name:
Chemiluminescence NO instrument (Weinheimer/Montzka/Tyndall)
What is being measured:
Nitric oxide (NO)
Sampling Protocol:
Online
Manufacturer:
Custom
Model:
Custom
Instrument year
:
None specified
Data recording software:
Inquire with PI
Data analysis software:
Inquire with PI
Raw data time resolution:
1 s
Analysis data averaging:
Inquire with PI
Detection limit:
25 pptv
Sensitivity to temperature (and correction method, if applicable):
:
Inquire with PI
Sensitivity to relative humidity (and correction method, if applicable):
:
Inquire with PI
Sampling method:
Direct sampling
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
Instrument flow rate:
Tubing inner diameter:
Tubing material:
PFA Teflon
Chemical identification method:
Chemiluminescence upon reaction with ozone and detection of light
Data analysis method:
Inquire with PI
Quantification method:
Applying NO calibration factor
Calibration method:
Direct calibration with NIST traceable standards
Calibration drift estimate:
Calibration schedule:
Uncertainty estimation method:
Inquire with PI, 10 pptv + 10% for NO (typical)
Known interferences:
Link to supplemental information:
Additional notes:
Measurement uncertainty:
Measurement units: