Instrument - High Resolution - Time of Flight - Aerosol Mass Spectrometer
Short name:
HR-ToF-AMS
Full name:
High Resolution - Time of Flight - Aerosol Mass Spectrometer
What is being measured:
Particle Elemental Composition & Fragments
Sampling Protocol:
Online
Manufacturer:
Aerodyne
Model:
HR-ToF-AMS
Instrument year
:
2005
Data recording software:
Built-in DAQ programs
Data analysis software:
Custom IGOR script (Squirrel)
Raw data time resolution:
1s
Analysis data averaging:
60 ~ 90s
Detection limit:
0.003 ug/m3 in V mode; 0.05 in W mode
Sensitivity to temperature (and correction method, if applicable):
:
N/A
Sensitivity to relative humidity (and correction method, if applicable):
:
N/A
Sampling method:
Direct gas sampling
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
Instrument flow rate:
0.08 lpm
Tubing inner diameter:
1/4'' OD; 0.152'' ID
Tubing material:
Copper
Chemical identification method:
Data analysis method:
The built-in DAQ program converts the raw data to formatted .HDF files, then those files are pulled out by a custom IGOR script and converted to its database.
Quantification method:
Calibration method:
Varied concentrations of atomized Ammonium Sulfate Particles were size-selected by a fixed-voltage SMPS and injected to AMS, then calibration curve could be built based on the real time SMPS counts and AMS mass readings.
Calibration drift estimate:
Low (on the order of weeks)
Calibration schedule:
Monthly
Uncertainty estimation method:
9999
Known interferences:
N/A
Link to supplemental information:
N/A
Additional notes:
N/A
Measurement uncertainty:
(a) Local_Time (N/A); (b) Experimental_Time (N/A); (c) O:C (+/- ?%); (d) H:C (+/- ?%); (e) f43 (+/- ?%); (f) f44 (+/- ?%)
Measurement units:
(a) Local_Time (N/A); (b) Experimental_Time (min); (c) O:C (N/A); (d) H:C (N/A); (e) f43 (N/A); (f) f44 (N/A)