Instrument - Model 49i UV Photometric O3 Analyzer
Short name:
Ozone analyzer
Full name:
Model 49i UV Photometric O3 Analyzer
What is being measured:
Ozone (UV photometry)
Sampling Protocol:
Online
Manufacturer:
Thermo Scientific
Model:
49i
Instrument year
:
2012
Data recording software:
MICAS-X
Data analysis software:
NA
Raw data time resolution:
1 sec
Analysis data averaging:
4 sec
Detection limit:
0.25ppb RMS (zero noise)
Sensitivity to temperature (and correction method, if applicable):
:
Unknown
Sensitivity to relative humidity (and correction method, if applicable):
:
Unknown
Sampling method:
Direct gas sampling through particle filter
Sample preparation method:
NA
Sample residence time (chamber to instrument) (seconds):
1
Length of tubing (cm):
Instrument flow rate:
Tubing inner diameter:
Tubing material:
FEP Teflon
Chemical identification method:
Data analysis method:
Quantification method:
Onboard instrument
Calibration method:
Factory calibration or comparison to NOx actinometry
Calibration drift estimate:
Low (on the order of weeks)
Calibration schedule:
Variable, as drift detected by NOx actinometry
Uncertainty estimation method:
Based on periodic comparison
Known interferences:
Alkenes
Link to supplemental information:
Additional notes:
Measurement uncertainty:
<10%
Measurement units:
ppb