Instrument - Scanning Mobility Particle Sizer - Size Distribution

Short name:
SMPS - SizeD

Full name:
Scanning Mobility Particle Sizer - Size Distribution

What is being measured:
Particle Size Distribution

Sampling Protocol:
Online

Manufacturer:
Home-made

Model:
TSI3081, TSI3771

Instrument year :
2005

Data recording software:
Labview Program

Data analysis software:
Custom Excel Program

Raw data time resolution:
1s

Analysis data averaging:
60 ~ 120s

Detection limit:
#: 0.108 cm-3; V: 0.001 um3/cm3

Sensitivity to temperature (and correction method, if applicable): :
N/A

Sensitivity to relative humidity (and correction method, if applicable): :
N/A

Sampling method:
Direct gas sampling

Sample preparation method:
None

Sample residence time (chamber to instrument) (seconds):

Length of tubing (cm):

Instrument flow rate:
Sheath 2.5 lpm; Main 0.25 lpm

Tubing inner diameter:
1/4'' OD; 0.152'' ID

Tubing material:
Copper

Chemical identification method:
None

Data analysis method:
The SMPS data that were recorded in scanning mode are analyzed using a custom Excel script. Particle size distribution, number/volume concentration as a function of time are displayed. Custom wall loss fitting time is available to do wall loss correction as needed.

Quantification method:
Signal strength is converted to partial number/size distribution by designed transfer functions.

Calibration method:
Standard particles(created by standard PSL solution) with a certain size range are sampled and detected. The calibration factor will be calculated based on the difference.

Calibration drift estimate:
No estimate

Calibration schedule:
Annually

Uncertainty estimation method:
Duplication of the same experimental conditions.

Known interferences:
N/A

Link to supplemental information:
N/A

Additional notes:
N/A

Measurement uncertainty:
(a) Local_Time (N/A); (b) Experimental_Time (N/A); (c) Bin_ab.cd (+/- ?%)

Measurement units:
(a) Local_Time (N/A); (b) Experimental_Time (min); (c) Bin_ab.cd (N/A)