Instrument - High Performance Liquid Chromatograph

Short name:
Agilent HPLC

Full name:
High Performance Liquid Chromatograph

What is being measured:
Organic particles, gas-phase organics, derivatized organic particles

Sampling Protocol:
Offline

Manufacturer:
Agilent

Model:
1100

Instrument year :
None specified

Data recording software:
Agilent Chemstation 2.0

Data analysis software:
Manual Data Analysis

Raw data time resolution:
2 s

Analysis data averaging:

Detection limit:
100 ng

Sensitivity to temperature (and correction method, if applicable): :
N/A

Sensitivity to relative humidity (and correction method, if applicable): :
N/A

Sampling method:
Injection of liquid extract

Sample preparation method:
Filters extracted in ethyl acetate

Sample residence time (chamber to instrument) (seconds):

Length of tubing (cm):

Instrument flow rate:

Tubing inner diameter:

Tubing material:
N/A

Chemical identification method:
UV-Vis absorption, retention time matching to standards.

Data analysis method:

Quantification method:
Peak areas are converted to concentration by plotting on a calibration curve of standards, either commercial or generated, the amount of solvent used to extract filters, the injection volume, and the mass of the filter, correcting for particle wall loss.

Calibration method:
Calibration curve of commercial, or fractionated standards. For fractionated standards, they are collected in preweighed vials, reconstituted in solvent and re-injected to determine senstivity.

Calibration drift estimate:
Low (on the order of weeks)

Calibration schedule:

Uncertainty estimation method:
Uncertainty is estimated at +/- 5%, and propagated from standard deviations of the calibration curve, uncertainties in injection volume.

Known interferences:

Link to supplemental information:

Additional notes:

Measurement uncertainty:
All measurements from this instrument have uncertainties of 5% of absolute value.

Measurement units:
All measurements from this instrument have units of mass.