Instrument - Temperature & humidity probes + ozone and NOx monitors
Short name:
T/RH/O3/NOx
Full name:
Temperature & humidity probes + ozone and NOx monitors
What is being measured:
temperature, relative humidity, O3, NO, NO2/NOy
Sampling Protocol:
Online
Manufacturer:
Thermo Scientific, Vaisala
Model:
42i, 49i
Instrument year
:
None specified
Data recording software:
Custom labview program
Data analysis software:
Raw data time resolution:
30 s
Analysis data averaging:
Detection limit:
1 ppb O3, 0.5 ppb NOx
Sensitivity to temperature (and correction method, if applicable):
:
Sensitivity to relative humidity (and correction method, if applicable):
:
Sampling method:
Direct sampling
Sample preparation method:
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
100
Instrument flow rate:
1 lpm
Tubing inner diameter:
0.16 in
Tubing material:
FEP Teflon
Chemical identification method:
O3: Photometric (UV absorption)
NO: Chemiluminescence
NO2: Conversion to NO by molybdenum catalyst, then chemiluminescence
Data analysis method:
Quantification method:
Calibration method:
NOx: calibrated with NO & NO2 primary standards
O3: cross-calibrated with an ozone generator & FT-IR spectrometer
RH: salt solutions
Calibration drift estimate:
Low (on the order of weeks)
Calibration schedule:
Annually
Uncertainty estimation method:
Inquire with PI
Known interferences:
NOy can interfere with NO2 by conversion on molybdenum catalyst; high (ppm) levels of H2O2 can interfere with O3
Link to supplemental information:
Additional notes:
Measurement uncertainty:
NOx: 20%; O3: 10%; RH/T: 10%
Measurement units:
T: ˚C; RH: %; O3/NOx: ppb.