Instrument - VOCUS CI-TOF
Short name:
VOCUS
Full name:
VOCUS CI-TOF
What is being measured:
VOC chemical speciation
Sampling Protocol:
Online
Manufacturer:
TOFWERK, Aerodyne Research, Inc
Model:
2R, with LTOF
Instrument year
:
circa 2019
Data recording software:
TofDaq, Acquility
Data analysis software:
Tofware
Raw data time resolution:
1 s
Analysis data averaging:
None
Detection limit:
sub ppt
Sensitivity to temperature (and correction method, if applicable):
:
Unknown. Not relevant for data posted since instrument was at stable laboratory temperatures.
Sensitivity to relative humidity (and correction method, if applicable):
:
No dependence (since high water mixing ratios in the IMR region by design)
Sampling method:
~1-2 liters per minute through 1/4 inch OD (~1/8 inch ID) PFA Teflon 10 m tubing
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
2.4
Length of tubing (cm):
1000
Instrument flow rate:
~0.1 lpm (sampling inlet flowrate to instrument ~2 lpm)
Tubing inner diameter:
0.32 cm
Tubing material:
PFA Teflon
Chemical identification method:
exact mass (m/z), aided by knowledge of chemical system under study and precursors added.
Data analysis method:
Standard high-resolution fitting using Tofware software in Igor Pro (Wavemetrics)
Quantification method:
Daily calibrations with multiple VOC from calibration cylinder and sometimes known injection quantity of VOC into chamber of known volume. Relationship between ions/s and mixing ratio established.
Calibration method:
See "quantification method". Calibration cylinder air mixed with zero air or as standard addition at several concentrations levels (controlled with automated routine).
Calibration drift estimate:
No estimate
Calibration schedule:
Prior to each chamber experiment
Uncertainty estimation method:
Accuracy uncertainty primarily from uncertainty in calibration. Precision uncertainty from ion counting statistics.
Known interferences:
Compounds with same or very similar m/z.
Link to supplemental information:
https://www.tofwerk.com/products/vocus/
https://pubs.acs.org/doi/full/10.1021/acs.analchem.8b02641
Additional notes:
Measurement uncertainty:
TBD
Measurement units:
ppb?