Instrument - Atmospheric pressure interface toluene chemical ionization mass spectrometer
Short name:
Toluene CIMS
Full name:
Atmospheric pressure interface toluene chemical ionization mass spectrometer
What is being measured:
Gas-phase siloxanes and oxidation products
Sampling Protocol:
Online
Manufacturer:
Tofwerk/Aerodyne
Model:
APi-LTOF-CIMS
Instrument year
:
2016
Data recording software:
Tofwerk's TOFDAQ
Data analysis software:
Tofwerk's TofWare and further processing with Igor scripts
Raw data time resolution:
1 Hz
Analysis data averaging:
Data is averaged to 1 minute resolution offline.
Detection limit:
3x the standard deviation of the noise
Sensitivity to temperature (and correction method, if applicable):
:
Sensitivity to relative humidity (and correction method, if applicable):
:
Sampling method:
Direct sampling with teflon tubing
Sample preparation method:
No sample preparation
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
1000
Instrument flow rate:
1.7 slpm
Tubing inner diameter:
3/16 in
Tubing material:
PFA Teflon
Chemical identification method:
m/z values are assigned chemical formulas based on high-resolution peak fitting with constrained isotopes. The identities of the chemicals are determined from what elements are expected to be present and how well the predicted isotopes match the measured distribution.
Data analysis method:
Quantification method:
Calibration method:
Calibration drift estimate:
Calibration schedule:
Uncertainty estimation method:
Uncertainty should follow counting error statistics when units are report in normalized counts per second unless otherwise specified. For calibrated compounds, uncertainty in the calibration is included via error propagation.
Known interferences:
Link to supplemental information:
Additional notes:
Measurement uncertainty:
Measurements have uncertainties equal to the square root of the counted ions, according to Poisson statistics.
Measurement units:
Normalized counts per second