Instrument - NO/NO2 Gas Analyzer
Short name:
NOx Box
Full name:
NO/NO2 Gas Analyzer
What is being measured:
NO is detected and NO2 is measured indirectly.
Sampling Protocol:
Online
Manufacturer:
Teledyne
Model:
T200UP
Instrument year
:
2019
Data recording software:
NumaView within the analyzer
Data analysis software:
Data is loaded into Igor for processing
Raw data time resolution:
0.1Hz
Analysis data averaging:
No
Detection limit:
50 ppt
Sensitivity to temperature (and correction method, if applicable):
:
Sensitivity to relative humidity (and correction method, if applicable):
:
Sampling method:
Direct sampling with teflon tubing
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
20
Instrument flow rate:
1 lpm
Tubing inner diameter:
3/16in
Tubing material:
PFA Teflon
Chemical identification method:
Data analysis method:
Quantification method:
Calibration method:
Calibration drift estimate:
Calibration schedule:
Uncertainty estimation method:
Uncertainty in the measurement is taken as the reported uncertainty from the manufacturer, 0.5%
Known interferences:
Certain NOy (e.g., peroxyacetylnitrate) are “read” as NO2 on the NOx analyzer. Can also have interferences from excited OH.
Link to supplemental information:
Additional notes:
Measurement uncertainty:
All measurements have 0.5% uncertainty
Measurement units:
ppb