Instrument - MAGIC water condensation particle counter with a custom diethylene glycol inlet
Short name:
MAGIC DEG wCPC
Full name:
MAGIC water condensation particle counter with a custom diethylene glycol inlet
What is being measured:
Aerosol particles down to 1-nm in size
Sampling Protocol:
Online
Manufacturer:
Aerosol Dynamics Inc.
Model:
Custom MAGIC
Instrument year
:
None specified
Data recording software:
Data analysis software:
Raw data time resolution:
1 hz
Analysis data averaging:
Detection limit:
~1 nm
Sensitivity to temperature (and correction method, if applicable):
:
Sensitivity to relative humidity (and correction method, if applicable):
:
Sampling method:
Sample preparation method:
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
10
Instrument flow rate:
0.3 LPM
Tubing inner diameter:
1.6 mm
Tubing material:
Stainless Steel
Chemical identification method:
Data analysis method:
Quantification method:
Calibration method:
Calibration performed using a half-mini DMA and electrometer
Calibration drift estimate:
Calibration schedule:
As Needed
Uncertainty estimation method:
Particle uncertainty could be estimated with the use of an half-mini DMA and electrometer sampling in-tandem with the CPC. This has been done previously to show the size cut-off for the CPC. Uncertainty will be higher at particle concentrations greater than 100,000.
Known interferences:
Link to supplemental information:
Additional notes:
Measurement uncertainty:
Uncertainty depends on particle size and concentration. Overall uncertainty estimated to be +/- 5%
Measurement units:
# cm-3