Instrument - HR-ToF-AMS
Field Value
Short name AMS (Far, time series)
Full name HR-ToF-AMS
What is being measured Non-refractory particle composition
Sampling Protocol Online
Manufacturer Aerodyne
Model N/A
Instrument year None specified
Data recording software ToF-AMS DAQ 4.024beta
Data analysis software SQ 1.62A, PIKA 1.22A
Raw data time resolution Every minute
Analysis data averaging
Detection limit 0.003 μg/m3
Sensitivity to temperature (and correction method, if applicable): Yes
Sensitivity to relative humidity (and correction method, if applicable): Yes
Sampling method Direct sampling
Sample preparation method
Sample residence time (chamber to instrument) (seconds)
Length of tubing (cm) 290
Instrument flow rate 0.08
Tubing inner diameter 1/4 inch
Tubing material Stainless Steel
Chemical identification method Mass spectrometery with electron impact ionization.
Data analysis method
Quantification method
Calibration method Ionization efficiency and relative ionization effeciencies of the instrument are calibrated with pure 300nm ammonium nitrate particles and pure 300nm ammonium sulfate particles every other week.
Calibration drift estimate
Calibration schedule As Needed
Uncertainty estimation method Filter runs before the sampling
Known interferences Air and inorganic salts can inteference with organic signals
Link to supplemental information
Additional notes AMS species concentration need to consider collection efficiency change during the experiment.
Measurement uncertainty 40%
Measurement units ug/m3