Short name |
AMS (Far, time series) |
Full name |
HR-ToF-AMS |
What is being measured |
Non-refractory particle composition |
Sampling Protocol |
Online |
Manufacturer |
Aerodyne |
Model |
N/A |
|
|
Instrument year
|
None specified |
Data recording software |
ToF-AMS DAQ 4.024beta |
Data analysis software |
SQ 1.62A, PIKA 1.22A |
Raw data time resolution |
Every minute |
Analysis data averaging |
|
Detection limit |
0.003 μg/m3 |
Sensitivity to temperature (and correction method, if applicable):
|
Yes
|
Sensitivity to relative humidity (and correction method, if applicable):
|
Yes
|
Sampling method |
Direct sampling |
Sample preparation method |
|
Sample residence time (chamber to instrument) (seconds) |
|
Length of tubing (cm) |
290 |
Instrument flow rate |
0.08 |
Tubing inner diameter |
1/4 inch |
Tubing material |
Stainless Steel |
Chemical identification method |
Mass spectrometery with electron impact ionization. |
Data analysis method |
|
Quantification method |
|
Calibration method |
Ionization efficiency and relative ionization effeciencies of the instrument are calibrated with pure 300nm ammonium nitrate particles and pure 300nm ammonium sulfate particles every other week. |
Calibration drift estimate |
|
Calibration schedule |
As Needed |
Uncertainty estimation method |
Filter runs before the sampling |
Known interferences |
Air and inorganic salts can inteference with organic signals |
Link to supplemental information |
|
Additional notes |
AMS species concentration need to consider collection efficiency change during the experiment. |
Measurement uncertainty |
40% |
Measurement units |
ug/m3 |