Particle Loss Characterization #30 - 2013-03-24
ID:
30
Name:
Particle Wallloss RH50
Instrument:
Soft X-ray Differential Mobility Analyzer in Negative Mode (#413)
Chamber:
Near/Far Bag 2012-2015 (Tran)
Temperature (°C):
22
Relative humidity (%):
50
Particle composition:
- Ammonium sulfate
Particle measurement size range:
16-915 nm
Particle generation method:
Atomization
Date of characterization experiment:
2013-03-24
Months since calibration:
Wall loss experiment description:
Introduce desired amounts of relative humidity (RH). Atomize particles into the bag. Mix briefly, and monitor wall loss over the course of several hours.
Wall loss rates calculation method:
Fit each particle diameter size bin with their loss rates (s-1)
Frequency of particle loss experiments:
Usually they are done for each experiment set but the rates don't change much throughout the years
Wall loss correction description:
At each particle size bin, the loss rate is used to calculate how many particles were lost and then added back on. We check that this correct doesn't increase or decrease particle volume during times when no reaction is occuring.
Link to supplemental information:
Additional notes:
Characterization data file:
130324_particlewallloss_RH50.csv