Experiment: UCR-CECERT-APL-Cocker Group_20150201_1-METHYLNAPHTHALENE/H2O2/M-XYLENE_Hydroxyl radical_No Seed_EPA2017A - Candice
Experiment
Name: UCR-CECERT-APL-Cocker Group_20150201_1-METHYLNAPHTHALENE/H2O2/M-XYLENE_Hydroxyl radical_No Seed_EPA2017A - Candice
Date: 2015-02-01
Organization
UCR-CECERT-APL-Cocker Group
Chamber
EPA chamber
Experiment Set
Sara Li_Old test
Goals
Photo-oxidation of two VOCs with OH under black lights.
Summary
Target: 1-methylnaphthalene 30 ppb+H2O2 1 ppm (both) +m-xylene 90 ppb(A) INSTRUMENTS: SMPS, GC1, AMS, APM, SYFTExperiment Details
Categories:
- Photolysis, any phase
- Aerosol formation
- Aerosol aging
Relative Humidity: 0.01
Seeded Experiment: No
Fuchs Adjusted Surface Area:
Condensation sink rate coefficient: 0.01
Reaction type:
Oxidant Name: Hydroxyl radical
Initial Oxidant Concentration Value: N/A (molecules_per_cubic_centimeter)
RO2 Main Fate: Not Sure
RO2 Lifetime: (Seconds)
Pressure: 0.015'' H2O
Data Sets
ID | Instrument | File | |
---|---|---|---|
3068 | SMPS |
EPA2017A_SMPS.csv
|
|
3069 | APM-SMPS |
EPA2017A_APM.csv
|
|
3070 | Ozone Box |
EPA2017A_O3.csv
|
|
3071 | NOx Box |
EPA2017A_NOx.csv
|
|
3072 | SMPS - SizeD |
EPA2017A_SizeD.csv
|
|
3073 | TC |
EPA2017A_Temperature.csv
|
|
3074 | Licor |
EPA2017A_Licor.csv
|
Data Set plots
Experiment Set Details
Name: Sara Li_Old test
Time format and units:
Description:
To investigate the SOA yield from DEGEE in low NOx condition.
Collaborators:
Contacts
Principal investigator:
David Cocker
E-mail: dcocker@engr.ucr.edu
Data Manager:
Chen Le
Phone: 9516603455
E-mail: cle030@ucr.edu
20150201_EPA2017_TimeLine.csv
Timeline
Time (America/Los_Angeles) | Action |
09:13 | NO span starts; flow 2|20|35|off |
09:30 | NO 88.38; NOx 89.36; CO 15.28; GPT span starts; flow 2|20|35|0.7 |
09:42 | Fill air off; Air Handlers on; Heaters on; Chiller #1 on |
09:44 | Inject 100 uL PFH into both bags |
09:45 | NO 62.99; NOx 88.38; CO 15.28; O3 span starts; flow 2|20|off|0.7 |
09:59 | Inject 30 uL 1-methylnaphthalene into both bags @45 V; 49 *C; 69 *C |
10:23 | Inject 400 uL H2O2 into both bags; 40 V; 69 *C; 10 lpm |
10:24 | O3 36.89; Starting E--W--Z sampling |
10:42 | Mix bags E-W for 10 mins |
10:55 | Close valve 14 |
10:59 | Inject 36 uL m-xylene into W bag |
11:10 | Mix W bag for 1 min |
11:32 | BL on; heaters off |
20:08 | BLs off; air handlers off; W 60 E 124 |
Citations
Characterizations
(None)
Particle Loss Characterizations:
(None)
Vapor Loss Characterizations:
(None)
Chamber Details
Name: EPA chamber
Volume in cubic meters: 84.2
Shape: Rectangular Prism
Surface area in cubic meters: 126
Flow Mode: Batch
Flow through rate in liters per minute:
Body Material: Teflon
Body thickness: 2
Replacement date: 2017-11-07
Replacement frequency: As needed
Cleaning method:
Flushing with air
Flushing with air with UV lights on
Oxidation
Wipe with ultra-pure water if needed
o3 background: 10
nox background: 0
Particle background in micrograms per cubic meter: 0
Air filtration method: The chamber is using zero air from our air purification system
Mixing method: 0
Mixing method material:
Temparature Control: Yes
Temparature measurements recorded: Yes
Temparature range (min degrees celcius): 5
Temparature range (max degrees celcius): 45
Humidity Control: Yes
Humidity Measurements Recorded: Yes
Relative humidity range (min percent): 0.01
Relative humidity range (max percent): 80
Particle experiment capability: 1
Lights: Yes
Light source type(s):
350nm UVA Black Lamps
Supplemental information:
Additional Notes: