Experiment: UCR-CECERT-APL-Cocker Group_20110421_ISOPRENE/H2O2_Hydroxyl radical_No Seed_EPA1417A - Chris
Experiment
Name: UCR-CECERT-APL-Cocker Group_20110421_ISOPRENE/H2O2_Hydroxyl radical_No Seed_EPA1417A - Chris
Date: 2011-04-21
Organization
UCR-CECERT-APL-Cocker Group
Chamber
EPA chamber
Experiment Set
Sara Li_Old test
Goals
Photo-oxidation of isoprene with OH under black lights
Summary
Targets: isoprene 0.250 ppm both H2O2 3 ppm West 1.5 ppm East INSTRUMENTS: GC, SMPS, AMS, VTDMA, KANOMAX, PILS-ToF
Additional Notes
East bag leakyExperiment Details
Categories:
- Photolysis, any phase
- Aerosol formation
- Aerosol aging
Relative Humidity: 0.01
Seeded Experiment: No
Fuchs Adjusted Surface Area:
Condensation sink rate coefficient: 0.01
Reaction type:
Oxidant Name: Hydroxyl radical
Initial Oxidant Concentration Value: N/A (molecules_per_cubic_centimeter)
RO2 Main Fate: HO2
RO2 Lifetime: (Seconds)
Pressure: 0.015'' H2O
Data Sets
ID | Instrument | File | |
---|---|---|---|
3221 | SMPS |
EPA1417A_SMPS.csv
|
|
3222 | Ozone Box |
EPA1417A_O3.csv
|
|
3223 | NOx Box |
EPA1417A_NOx.csv
|
|
3224 | SMPS - SizeD |
EPA1417A_SizeD.csv
|
|
3225 | TC |
EPA1417A_Temperature.csv
|
Data Set plots
Experiment Set Details
Name: Sara Li_Old test
Time format and units:
Description:
To investigate the SOA yield from DEGEE in low NOx condition.
Collaborators:
Contacts
Principal investigator:
David Cocker
E-mail: dcocker@engr.ucr.edu
Data Manager:
Chen Le
Phone: 9516603455
E-mail: cle030@ucr.edu
Reactants
Reactant | Compound type | Concentration | Concentration units |
---|---|---|---|
ISOPRENE | 250 | Parts Per Billion | |
H2O2 | 3 | Parts Per Million |
20110421_EPA1417_TimeLine.csv
Timeline
Time (America/Los_Angeles) | Action |
07:27 | NO/CO span flow 2|40|40|off |
07:52 | NO 99.95; NOx 98.34; CO 10.16 |
07:53 | GPT span flow 2|40|2|40 |
08:03 | NO 22.80; NOx 91.02 |
08:04 | O3 span flow 2|40|2|off |
08:14 | O3 #2 82.03 |
08:15 | Start E-W-Z sampling |
08:33 | Turned off fill air; turned on air handler |
08:41 | Inject 100 uL perfluorohexan into both bags |
08:44 | Inject 172.5 uL isoprene into both bags |
08:56 | Mix E-W for 7 mins |
09:13 | Closed valve 14 |
09:16 | Inject 300 uL 50% H2O2 into E |
09:29 | Inject 400 uL 50% H2O2 into W |
09:40 | Injected 200 uL 50% H2O2 into A rtr |
09:53 | All blacklights on |
16:36 | lights off |
Citations
Characterizations
(None)
Particle Loss Characterizations:
(None)
Vapor Loss Characterizations:
(None)
Chamber Details
Name: EPA chamber
Volume in cubic meters: 84.2
Shape: Rectangular Prism
Surface area in cubic meters: 126
Flow Mode: Batch
Flow through rate in liters per minute:
Body Material: Teflon
Body thickness: 2
Replacement date: 2017-11-07
Replacement frequency: As needed
Cleaning method:
Flushing with air
Flushing with air with UV lights on
Oxidation
Wipe with ultra-pure water if needed
o3 background: 10
nox background: 0
Particle background in micrograms per cubic meter: 0
Air filtration method: The chamber is using zero air from our air purification system
Mixing method: 0
Mixing method material:
Temparature Control: Yes
Temparature measurements recorded: Yes
Temparature range (min degrees celcius): 5
Temparature range (max degrees celcius): 45
Humidity Control: Yes
Humidity Measurements Recorded: Yes
Relative humidity range (min percent): 0.01
Relative humidity range (max percent): 80
Particle experiment capability: 1
Lights: Yes
Light source type(s):
350nm UVA Black Lamps
Supplemental information:
Additional Notes: