Instrument - Scanning Mobility Particle Sizer Particle Loss
Short name:
SMPS particle loss
Full name:
Scanning Mobility Particle Sizer Particle Loss
What is being measured:
Particle wall loss first-order rates
Sampling Protocol:
Online
Manufacturer:
Custom
Model:
Custom
Instrument year
:
None specified
Data recording software:
Labview
Data analysis software:
Matlab
Raw data time resolution:
N/A
Analysis data averaging:
N/A
Detection limit:
1E-4 s-1
Sensitivity to temperature (and correction method, if applicable):
:
N/A
Sensitivity to relative humidity (and correction method, if applicable):
:
N/A
Sampling method:
Direct sampling
Sample preparation method:
None
Sample residence time (chamber to instrument) (seconds):
Length of tubing (cm):
30
Instrument flow rate:
1 LPM
Tubing inner diameter:
Tubing material:
Stainless Steel
Chemical identification method:
Not applicable - no chemicals detected
Data analysis method:
The number concentration is graphed against time at each particle size bin. An exponential fit is done to obtain the first-order loss rate at each size bin.
Quantification method:
A custom inversion protocol in Matlab is done to obtain the particle's diameter.
Calibration method:
Size calibration with polystyrene latex spheres.
Calibration drift estimate:
Calibration schedule:
Uncertainty estimation method:
Inquire with PI
Known interferences:
Link to supplemental information:
Additional notes:
Measurement uncertainty:
30%
Measurement units:
Particle diameter (nm), Loss rate (s-1)
Characterizations
2013-03-22 (#9)
2013-03-24 (#10)
2013-03-26 (#11)
2013-03-06 (#12)
2013-03-07 (#13)
2013-03-14 (#14)
2013-03-09 (#15)
2013-03-11 (#16)