Chamber: Seinfeld chambers

Seinfeld chambers

Name: Seinfeld chambers

Organization
2014 FIXCIT Study

Volume in cubic meters: 22
Shape: Rectangular Prism
Surface area in cubic meters: 48
Flow Mode: Batch
Flow through rate in liters per minute:
Body Material: Teflon
Body thickness: 2
Replacement date:
Replacement frequency: As needed
Cleaning method: Flushing with air Flushing with air with UV lights on Oxidation
o3 background: < 0.5 ppb
nox background: 40 ppt
Particle background in micrograms per cubic meter: < 0.01 ug/m3
Air filtration method: Custom filtration canisters to remove NOx, hydrocarbons, water, particles, ozone, SOx from pressurized ambient air.
Mixing method: 1
Mixing method material: Pulsed air jets, manually actuated
Temparature Control: Yes
Temparature measurements recorded: Yes
Temparature range (min degrees celcius): 10
Temparature range (max degrees celcius): 45
Humidity Control: Yes
Humidity Measurements Recorded: Yes
Relative humidity range (min percent): 4
Relative humidity range (max percent): 90
Particle experiment capability: 1
Lights: Yes
Light source type(s): UV black lamps (350 nm centered)
Supplemental information: https://www.worldscientific.com/doi/abs/10.1142/9789813147355_0001
Additional Notes:


Experiment Sets (1)

ID Name Description Experiments (count)
162 FIXCIT experiments The Focused Isoprene eXperimen... 26